BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Safety Notice for Fixation Posts - Leksell Stereotactic System® including Leksell Gamma Knife®, Elekta

Product group Radiotherapy / Radiation protection
Reference 03345/09

Elekta has issued a Field Safety Notice concerning a problem with the Leksell Stereotactic System®, including the user of Leksell Gamma Knife®. Elekta has received some reports of heat generated at the tip of the fixation screws when used with the standard posts.

Safety Notice for Fixation Posts - Leksell Stereotactic System® including Leksell Gamma Knife®, Elekta , Download_VeroeffentlichtAm_EN PDF, 145KB, File does not meet accessibility standards