BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Safety Notice concerning a problem with the digital linear accelerators of type ARTISTE™ and ONCOR™ equipped with 160 MLC™, Siemens Healthcare

Product group Radiotherapy / Radiation protection
Reference 01938/11

Siemens Healthcare has issued a field safety notice concerning a problem with the digital linear accelerators of type ARTISTE™ and ONCOR™ equipped with 160 MLC™. An issue has been reported, where the actual field shape of the MLC deviated from the originally prescribed field shape after operating the `Motion Stop` button followed by `reset`on the Control console several times within one beam.

Safety Notice concerning a problem with the digital linear accelerators of type ARTISTE™ and ONCOR™ equipped with 160 MLC™, Siemens Healthcare , Download_VeroeffentlichtAm_EN PDF, 27KB, File does not meet accessibility standards