BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Safety Notice for the Optical Guidance Platform (OPG) Software, Varian medical systems

Product group Radiotherapy / Radiation protection
Reference 02901/11

Varian medical systems has issued a field safety notice concerning an anomaly that has been identified with the Optical Guidance Platform (OPG) Software v2.6 and v2.6.1 where the transfer of datasets from treatment planning systems other than the FastPlan system result in a lateral offset error of up to 1mm.

Safety Notice for the Optical Guidance Platform (OPG) Software, Varian medical systems , Download_VeroeffentlichtAm_EN PDF, 50KB, File does not meet accessibility standards