BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Safety Notice concerning the Eclipse treatment planning system, Varian medical systems

Product group Radiotherapy / Radiation protection
Reference 03779/11

Varian medical systems has issued a field safety notice concerning an anomaly in Eclipse versions 8.0 to 8.6 whereby a particular plan approval workflow may result in a difference between the intended SSD value and the SSD value communicated to the treatment delivery system.

Safety Notice concerning the Eclipse treatment planning system, Varian medical systems , Download_VeroeffentlichtAm_EN PDF, 344KB, File does not meet accessibility standards