Anomaly identified with Eclipse Treatment Planning System, Varian medical systems
2012.03.28
Product group Radiotherapy / Radiation protection - radiotherapy treatment planning systems
Reference 01663/12
Varian medical systems has issued a field safety notice concerning an anomaly that has been identified with the Eclipse Treatment Planning System where the calculated dose distribution is not updated or invalidated when a recalculation or reload is performed while the compensator or fluence editor is open. This may go unnoticed.