Safety Notice concerning shifted Dose for XiO Proton Spot Beams, Elekta

Product group Radiotherapy / Radiation protection - radiotherapy treatment planning systems
Reference 03756/12

Elekta has issued a field safety notice concerning shifted Dose for XiO Proton Spot Beams. Mass densities, as seen by the Proton Spot algorithm, are shifted by one CT slice thickness in the superior direction. The resulting doses are therefore calculated on a shifted grid.

Safety Notice concerning shifted Dose for XiO Proton Spot Beams, Elekta , Download_VeroeffentlichtAm_EN PDF, 119KB, File does not meet accessibility standards