Corrective action of Focal and Monaco treatment planning software, Elekta Inc. IMPAC Medical Systems, Inc.

Product group Radiotherapy / Radiation protection - radiotherapy treatment planning systems
Reference 04421/12

Elekta Inc. IMPAC Medical Systems, Inc. has issued a Field Safety Notice concerning a corrective action of Focal (versions 4.64.00 – 4.70.00) and Monaco (versions 3.10.00 – 3.20.00) treatment planning software. If the “DICOM coordinates” option is enabled for reports incorrect shift coordinates are generated on the print out of the “Setup reference” dialog. This can result in incorrect patient positioning.