Safety Notice for the Eclipse Treatment Planning System, Varian medical systems
2012.10.15
Product group Radiotherapy / Radiation protection - radiotherapy treatment planning systems
Reference 06038/12
Varian medical systems has issued a Field Safety Notice concerning a problem with Eclipse Treatment Planning System where an outdated assigned Hounsfield Unit (HU) value can be used for dose calculation when the clock on the Eclipse Client Workstation is not synchronized with the Database/System Server.