BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Urgent Safety Notice for the Eclipse™ Treatment Planning System, Varian Medical Systems

Product group Radiotherapy / Radiation protection - radiotherapy treatment planning systems
Reference 00530/13

Varian Medical Systems has issued a Field Saftey Notice concerning an anomaly that has been identified with the Eclipse Treatment Planning System where the calculation of dose dynamic plans containing different energy fields is incorrect when the Pencil Beam Convolution 11.0 algorithm (PBC 11.0) is used.

Urgent Safety Notice for the Eclipse™ Treatment Planning System, Varian Medical Systems , Download_VeroeffentlichtAm_EN PDF, 2MB, File does not meet accessibility standards