Field Safety Notice concerning an issue with the dose volume statistics calculation in COMPASS 3.0, IBA Dosimetrie GmbH

Product group Radiotherapy / Radiation protection - nuclear medical therapeutic equipment
Reference 05423/13

IBA Dosimetrie GmbH has issued a Field Safety Notice concerning an issue with the dose volume statistics calculation in COMPASS 3.0. Under certain conditions, resampling of regions of interest (ROI) geometries to the dose grid representation is not correctly calculated and an additional voxel layer is added incorrectly on the superior side of each ROI.