BfArM - Federal Institute for Drugs and Medical Devices

Navigation and service

Safety Notice concerning Eclipse™ Treatment Planning System, ARIA® Radiation Oncology, ARIA®Radiation Therapy Management, and Acuity™, Varian medical systems

Product group Radiotherapy / Radiation protection - radiotherapy treatment planning systems
Reference 07371/13

Varian medical systems has issued a Field Safety Notice concerning an anomaly that has been identified with the Plan Parameters workspace within Eclipse™ Treatment Planning System, ARIA® Radiation Oncology, ARIA®Radiation Therapy Management, and Acuity™. When a dose relevant parameter is changed and “Do not clear MUs and Reference Point dose” is selected, the Monitor Units (MUs) may change.

Safety Notice concerning Eclipse™ Treatment Planning System, ARIA® Radiation Oncology, ARIA®Radiation Therapy Management, and Acuity™, Varian medical systems , Download_VeroeffentlichtAm_EN PDF, 32KB, File does not meet accessibility standards