Safety Notice for the product Desktop Pro™, Elekta
2014.01.23
Product group Radiotherapy / Radiation protection - radiation facilites with accelerators for teletherapy
Reference 07710/13
Elekta has issued a Field Safety Notice concerning recommended quality assurance after a calibration procedure of Desktop Pro™ R4.2, R5.0, R5.1, R6.0, R6.1. During the treatment table, gantry, and collimator calibration procedures, it is possible to type incorrect values. If you do not do the quality assurance (QA) procedures after the calibration, it is possible that you will not find the error made during the calibration.
Safety Notice for the product Desktop Pro™, Elekta , Download_VeroeffentlichtAm_EN