Safety Notice for the product Integrity™, Elekta

Product group Radiotherapy / Radiation protection - radiation facilites with accelerators for teletherapy
Reference 07712/13

Elekta has issued a Field Safety Notice concerning recommended quality assurance after a calibration procedure of Integrity™ R1.1, R3.0, and R3.1. During the treatment table, gantry, and collimator calibration procedures, it is possible to type incorrect values. If you do not do the quality assurance (QA) procedures after the calibration, it is possible that you will not find the error made during the calibration.