Field Safety Notice for Image-Com Software, TomTec Imaging Systems GmbH
2010.08.20
Product group Ultrasonics
Reference 02733/10
TomTec Imaging Systems GmbH has issued a Field Safety Notice about a calculation issue occurring in the Image-Com Measurement Module. The formula for the calculation of mean pressure gradient is not following the guidelines published by the ASE.