Urgent Safety Notice for the endocavitary SE12-3 transducer, SuperSonic Imagine
2013.03.14
Product group Ultrasonics - ultrasonic diagnostic equipment
Reference 00661/13
SuperSonic Imagine informed about missing information within the user manual of Aixplorer Ultrasound system when endocavitary SE12-3 transducer is used for the brain. In this case, the transducer must be used with an intraoperative protective cover.