Recall of Custom Sterile Packs containing BD Plastipak™ 50ml Luer Lok Syringes, Medline
2013.06.24
Product group Dressings / Pads - others
Reference 02959/13
Medline International Germany GmbH recalls Custom Sterile Packs containing BD Plastipak™ 50ml Luer Lok Syringes (ref. 300223 und 300865) with old design stopper. If syringes with old design stopper are used in syringe pumps that have been adjusted for the new stopper, there may be a delay in initiating an occlusion alarm.